**Background & Methods**
This analytical method development study created and validated a rapid UPLC-MS/MS (Ultra-Performance Liquid Chromatography-Tandem Mass Spectrometry) technique for simultaneous quantification of 39 cannabinoids and their metabolites in standard solutions. Chromatographic and mass-spectrometric conditions were systematically optimized, with performance evaluated against ANSI/ASB Standard 036 criteria including calibration performance, bias, precision, carryover, detection capability, and chromatographic selectivity.
**Key Findings**
• The method achieved excellent calibration performance for 38 of 39 analytes, with coefficients of determination (R²) ≥ 0.990. Only 7-OH-CBD showed lower performance (R² = 0.975) with a limit of quantification (LOQ) of 0.500 ng/mL, but was retained as a provisionally characterized analyte due to acceptable bias and precision despite inferior calibration metrics.
• Most critical isomeric pairs achieved baseline chromatographic separation, with the notable exception of Δ⁹-THCV/Δ⁸-THCV, which showed only partial resolution (Rs = 1.12), requiring further optimization for complete distinction.
• The method establishes a broad-panel quantification platform for major cannabinoids (THC, CBD) and 37 related compounds and metabolites, substantially expanding beyond existing methods that typically target only THC and CBD with limited related analytes.
**Dosage & Administration**
Not reported.
**Safety & Adverse Effects**
Not reported.
**Evidence Quality**
This is a methodological review/validation study limited to standard solutions without clinical application to biological, forensic, wastewater, or environmental matrices. The authors explicitly acknowledge that matrix effects, extraction recovery, stability, dilution integrity, and internal-standard correction require evaluation before the method can be reliably applied to real-world samples. The work represents preliminary analytical development rather than clinical evidence, with the main limitation being the lack of validation in actual biological or environmental samples. The partial resolution of one isomeric pair (Δ⁹-THCV/Δ⁸-THCV) suggests further optimization is needed. As a methodological study published in 2026, it contributes to analytical toxicology infrastructure but provides no direct clinical or safety data regarding cannabinoid use.